Portable High Range Conductivity & TDS

£242.80

SKU: c22e20e25087 Categories: ,

Description

Portable EC/TDS/Temperature Meter

The HI-99301 is a portable, EC/TDS/temperature meter that measures high range conductivity in mS/cm and TDS in ppt. Both instruments are housed in a case rated for IP67 conditions. 

This instrument offers single point, automatic calibration with automatically temperature compensated measurements. The compensation coefficient and EC/TDS conversion factor are user-selectable from 0.45 to 1.00 and temperature coefficient is selectable from 0.0 to 2.4%/°C. 

This meter easily fits in the palm of your hand and the bottom probe connection ensures the electrode cable doesn’t get in your way. The large, multi-level LCD displays the primary reading, temperature and calibration guides simultaneously. Symbols and messages on the LCD indicate meter status and guides users through operations. 

At start-up, the meter shows the remaining battery percentage and when a low battery condition is detected, a battery symbol appears on the LCD to advise the user that only a few hours of working time is left. 

The HI-76306 EC/TDS/temperature probe resists clogging and is easy to clean. 

Ideal for

  • Boiler Water
  • Waste Water

Main benefits

  • BEPS (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
  • Battery % displayed on start-up
  • Compact, heavy-duty and water resistant

Key features

  • Amperometric probe technology
  • Automatic Temperature Compensation
  • Automatic single calibration with calibration indicator
  • On-screen tutorial messages for calibration and set-up

Supplied with

  • HI-76306 EC/TDS probe
  • Batteries
  • Instructions and rugged carrying case

Specifications

Range EC 0.00 to 20.00 mS/cm
Range TDS 0.00 to 10.00 ppt (g/L)
Range Temperature 0.0 to 60.0°C / 32.0 to 140.0°F
Resolution EC 0.01 mS/cm
Resolution TDS 0.01 ppt (g/L)
Resolution Temperature 0.1°C / 0.1°F
Accuracy EC ±2% F.S.
Accuracy TDS ±2% F.S.
Accuracy Temperature ±0.5°C / ±1°F
TDS Factor adjustable, from 0.45 to 1.00 with 0.01 step
(default 0.50)
Calibration automatic, 1 point at 6.44 ppt (CONV 0.5)
or at 9.02 ppt (CONV 0.7)
or at 12880 µS/cm (others)
Temperature Compensation Automatic, 0 to 60°C (32 to 140°F) with ß adjustable
from 0.0 to 2.4% with 0.1% step
Probe HI 76306, EC/TDS/temperature, with DIN connector,
1 m (3.3′) cable (included)
Battery Type / Life 4 x 1.5V AAA / approx. 500 hours of continuous use
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 143 x 80 x 38 mm (5.6 x 3.1 x 1.5″)
Weight 210 g (7.4 oz.)